• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 BARD MESH PRE-SHAPED W/ KEYHOLE; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 BARD MESH PRE-SHAPED W/ KEYHOLE; SURGICAL MESH Back to Search Results
Catalog Number 0113710
Medical Device Problem Codes Defective Device (2588); Material Deformation (2976); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Hernia (2240); Injury (2348)
Date of Event 07/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Should additional information be provided, a supplemental emdr will be submitted.Not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard mesh large, pre-shaped with keyhole on (b)(6) 2015.As reported, the patient is making a claim for an adverse patient outcome against the bard mesh large, pre-shaped with keyhole.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the device was defective.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-apr-2010) is considered to be a best estimate.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard mesh large, pre-shaped with keyhole on (b)(6) 2015.As reported, the patient is making a claim for an adverse patient outcome against the bard mesh large, pre-shaped with keyhole.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the device was defective.Addendum per additional information provided: (b)(6) 2015 - patient was diagnosed with left inguinal hernia thereby underwent open repair with implant of bard pre-shaped mesh with keyhole.Per op notes, ¿a large direct hernia sac was noted and reduced.A bard pre-shaped mesh with keyhole was placed and tacked using sutures.¿ (b)(6) 2020 - patient was diagnosed with recurrent left inguinal hernia thereby underwent open repair with excision of bard pre-shaped mesh with keyhole and implant of synthetic mesh.Per op notes, ¿the prior mesh appeared to be moderate balled up was excised entirely.The hernia defect with cord lipoma was noted and excised.A synthetic mesh was placed and sutured.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD MESH PRE-SHAPED W/ KEYHOLE
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key10939604
Report Number1213643-2020-10797
Device Sequence Number14295120
Product Code FTL
UDI-Device Identifier00801741030703
UDI-Public(01)00801741030703(17)150428(10)HUUD0967
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/28/2015
Device Catalogue Number0113710
Device Lot NumberHUUD0967
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2020
Supplement Date Received by Manufacturer09/23/2025
Initial Report FDA Received Date12/03/2020
Supplement Report FDA Received Date10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/15/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexMale
Patient Weight90 KG
-
-