Model Number 72204402
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/09/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during an acl procedure, the biosure screw broke when it was being threaded into the patient.60% of the screw was removed with a grasper and the other distal 40% remained on the patient.The procedure was successfully completed without significant delay using a back-up device.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10 h3, h6 the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection performed on the product observed debris in the threads and the bottom half of the screw is broken off.Based on the condition of the product material found during visual inspection it was determined that additional material testing is not required.A review of the raw materials found that the storage requirements, material specifications, and material tests were appropriately documented.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of the complaint revealed there were no patient injuries reported and no apparent patient impact based on the details provided.Therefore, no further medical assessment is warranted.The complaint was confirmed.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device tip, attempted correction of a damaged device, or an inadvertent impact event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW BIOSURE REGENESORB 8MM X 35MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10939683
Report Number1219602-2020-01979
Device Sequence Number2419057
Product Code MAI
UDI-Device Identifier00885554036787
UDI-Public00885554036787
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K142948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/03/2023
Device Model Number72204402
Device Catalogue Number72204402
Device Lot Number50876995
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2021
Initial Date Received by Manufacturer 11/09/2020
Supplement Date Received by Manufacturer05/26/2021
Initial Report FDA Received Date12/03/2020
Supplement Report FDA Received Date05/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-