OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/09/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user that during the preparation for use, the front panel of the subject device intermittently froze.Other detailed information was not provided.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The device was returned to an olympus repair center for evaluation.The reported issue (front panel of the subject device intermittently froze) was found to be caused by a failure of the cp board (printed circuit board).Based on the legal manufacturer's investigation, since the device was manufactured over 10 years ago, it is likely that the parts on the cp board deteriorated over time due to long-term repeated use, causing the printed circuit board failure.The cp board controls the front panel; therefore, it is likely that the front panel froze intermittently due to the failure of the cp board.A review of the device history record (dhr) found no issues that could have caused or contributed to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10940176
MDR Text Key219561337
Report Number8010047-2020-09775
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/09/2020
Initial Date FDA Received12/03/2020
Supplement Dates Manufacturer Received01/15/2021
Supplement Dates FDA Received01/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-