| Model Number |
E4115 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the serial number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 02/2045).
|
| |
|
Event or Problem Description
|
|
It was reported that during a vacuum assisted breast biopsy procedure, the device allegedly had unintended cutting sequence inside the patient.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during a vacuum assisted breast biopsy procedure, the device allegedly had unintended cutting sequence inside the patient.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the event was determined to be expected and the investigation did not identify any manufacturing and/or service-related issues, therefore a device history record and a manufacturing review are not required.Investigation summary: the device was received for evaluation.The enspire was visually inspected upon receipt and was found to be in good condition.The device was tested and investigation has been determined to be unconfirmed for the reported device unintended cutting sequence inside the patient.The root cause cannot be determined as the problem could not be reproduced.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 02/2045).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|