According to the reporter, during a mastectomy, the electrode caught fire.According to medical staff, it burned like a small torch.They changed to another electrode to complete the procedure.There was no patient injury.
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found that the electrode had eschar on it, the coating was missing, and the packaging was not damaged.The electrode should be cleaned often to mitigate eschar build up.The eschar poses an ignition source and cause flames.It was reported that the device caught fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the sparking and heating associated with electrosurgery can provide an ignition source.Observe fire precautions at all times: when using electrosurgery in the same room with gases or flammable substances, prevent pooling of fluids and the accumulation of gases under surgical drapes or near the surgical site.Tissue buildup (eschar) on the tip of an active electrode poses a fire hazard, especially in oxygen-enriched environments, such as in throat or mouth procedures.Eschar plus high oxygen may create embers.Keep the electrode clean and free of all debris.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|