MEDOS INTERNATIONAL SÃ RL CH 5.5 EXP VERSE SCREW 6.0 X 40; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
|
Back to Search Results |
|
Model Number 199721640 |
Device Problem
Device-Device Incompatibility (2919)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional product code: kwp; kwq; mnh; mni; osh.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is jnj representative.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from (b)(6) reports an event as follows: it was reported that on (b)(6) 2020 the patient underwent for an unknown surgery.During the surgery, the screw has inner thread damaged due to multiple trying to engage correction key while the correction key has the inner grey part isn¿t locked.It was unknown if the surgery completed successfully.The patient outcome unknown.This complaint involves seven (7) devices.This report is for (1) 5.5 exp verse screw 6.0 x 40.This report is 1 of 7 for (b)(4).
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d11: concomitant product added to complaint.Investigation summary background: 1.Screw damage due to correction key damage 2.Inner grey part isn't locked.11/11/2020 updated event description: this complaint involves seven (7) devices.11/19/2020 updated event description: the screw has inner thread damaged due to multiple trying to engage correction key while the correction key has the inner grey part isn¿t locked.Investigation flow: device interaction/functional.Visual inspection: the 5.5 exp verse screw 6.0 x 4.0 (p/n: 199721640, lot #: 224314) was returned and received at us cq.Upon visual inspection, it was observed that the head of the screw was locked in the monoaxial position.Per the system guide dsus/spn/0115/0747(1), the polyaxial screw converts to a monoaxial screw upon tightening of the set screw.The system guide indicates that exceeding the system torque may result in failure of the implant.Upon loosening of the set screw, the screw head should become mobile again.Functional test: a functional assessment was performed on the complaint device.The head of the screw was unable to be moved from the received position.The returned verse correction keys (p/n: 199721000, lot #: 206341) was able to assemble to the verse screw.The verse correction key is being investigated in a different (b)(4).Can the complaint can be replicated with the returned device.Dimensional inspection: the outer diameter of the screw head was measured to be within the specification.Document/specification review: based on the date of manufacture, the current and manufactured revision of drawings were reviewed.Dwg-887046545, rev b/a.Dwg-887042533, rev: a.Dsus/spn/0115/0747(1).No design issues or discrepancies were identified.Complaint was confirmed.Investigation conclusion: the complaint condition is confirmed for the 5.5 exp verse screw 6.0 x 4.0 (p/n: 199721640, lot #: 224314).There is no indication that a design or manufacturing issue has caused the complaint condition and the possible root cause is over-torqueing of the set screws, however no definitive root cause could be identified.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h4, h6- the complaint condition is confirmed for the 5.5 exp verse screw 6.0 x 4.0 (p/n: 199721640, lot #: 224314).There is no indication that a design or manufacturing issue has caused the complaint condition and the possible root cause is over-torqueing of the set screws, however no definitive root cause could be identified.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: the dhr of product code: 199721640.Lot : 224314.Was electronically reviewed and the non-conformance nr-0112669.Has been observed during the manufacturing process, sorting done as part of the nr guaranty the product quality.The product was released on: 26.11.2018.Device history batch: null.Device history review: null.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|