• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD INTEGRA SYRINGE WITH DETACHABLE NEEDLE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD INTEGRA SYRINGE WITH DETACHABLE NEEDLE; PISTON SYRINGE Back to Search Results
Model Number 305272
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 11/08/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medical device expiration date: unknown.The customer's address is unknown.Device manufacture date: unknown.(b)(4).Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.The lot number is unknown, therefore device history record review (dhr) or quality notification review (qn) could not be performed.Investigation conclusion: unconfirmed: bd was not able to duplicate or confirm the customers indicated failure as no samples or photos were returned.Root cause description: no root cause can be determined as no samples were received.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that the detachable needle broke off of the bd integra" syringe during use and remained in the patient's buttocks.The patient went to the er and had an x-ray performed, but the broken needle was not found.The following information was provided by the initial reporter: "consumer reported --caller stated used integra syringe with needle break off in the buttocks.Consumer went to er (emergency room).The er completed an xray and could not locate the needle.They then took the syringe apart to look for the needle in syringe and still could not find it.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTEGRA SYRINGE WITH DETACHABLE NEEDLE
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
MDR Report Key10946794
Report Number1213809-2020-00871
Device Sequence Number3043639
Product Code MEG
UDI-Device Identifier30382903052722
UDI-Public30382903052722
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K011103
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number305272
Device Catalogue Number305272
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2020
Initial Report FDA Received Date12/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-