ABBOTT GMBH ALINITY C TOTAL BILIRUBIN REAGENT KIT; DIAZO COLORIMETRY, BILIRUBIN Back to Search Results
Model Number 04V5131
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.All available patient information is included.Additional patient details are not available.
 
Event or Problem Description
The customer reported a false decreased alinity c total bilirubin result for one patient.The customer provided the clinician questioned the initial result for (b)(6) of 51umol/l as it did not fit with the patient history.The repeat of 300umol/l does fit the patient history.There was no reported impact to patient management.
 
Additional Manufacturer Narrative
The evaluation of the customer¿s issue included a search for similar complaints, trending review, and labeling review.A review of tickets determined that there is normal complaint activity for alinity c total bilirubin (ln 04v51-31) lot 57727uq06.Trending review determined no trend for falsely depressed results for the product.Return testing was not completed as returns were not available.A review of the product labeling concluded that the issue is sufficiently addressed.Based on all available information no systemic issue or product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY C TOTAL BILIRUBIN REAGENT KIT
Common Device Name
DIAZO COLORIMETRY, BILIRUBIN
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
MDR Report Key10946833
Report Number3002809144-2020-01158
Device Sequence Number2177888
Product Code CIG
UDI-Device Identifier00380740158422
UDI-Public00380740158422
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K121985
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model Number04V5131
Device Catalogue Number04V51-31
Device Lot Number57727UQ06
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/18/2020
Supplement Date Received by Manufacturer01/18/2021
Initial Report FDA Received Date12/04/2020
Supplement Report FDA Received Date01/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY C PROCESSING MODU, 03R67-01, AC01970; ALNTY C PROCESSING MODU, 03R67-01, AC01970
-
-