| Model Number |
04V5131 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.All available patient information is included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer reported a false decreased alinity c total bilirubin result for one patient.The customer provided the clinician questioned the initial result for (b)(6) of 51umol/l as it did not fit with the patient history.The repeat of 300umol/l does fit the patient history.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of the customer¿s issue included a search for similar complaints, trending review, and labeling review.A review of tickets determined that there is normal complaint activity for alinity c total bilirubin (ln 04v51-31) lot 57727uq06.Trending review determined no trend for falsely depressed results for the product.Return testing was not completed as returns were not available.A review of the product labeling concluded that the issue is sufficiently addressed.Based on all available information no systemic issue or product deficiency was identified.
|
| |
|