| Model Number |
100/897/080CZ |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/21/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(6).Device evaluation in progress.
|
| |
|
Event or Problem Description
|
|
It was reported that the cuff on the tracheostomy tube ruptured.The product problem was observed immediately upon use.No patient injury or complications were reported in relation to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: returned device was received in used physical condition.During the evaluation of the device, there was a puncture in the inflation lumen of the tracheostomy.Such a defect can happen only if the adjustable flange is over pulled above the marking on the tube which is in conflict with the instructions for use.This was found to be caused by use error.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.
|
| |
|