• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DR. JAPAN CO., LTD. RELI; ANESTHESIA NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DR. JAPAN CO., LTD. RELI; ANESTHESIA NEEDLE Back to Search Results
Model Number SN22G351
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We are moving to close this complaint based on the manufacturer's investigation results.Although the suspect device was not returned by the end-user for examination, 20 samples from the same sterile lot were sent from retains for the factory to examine and test.Physical testing of the returned samples was performed for pull-out strength of the needle to needle hub bond, and the test samples met all specifications.In addition to the physical testing, a review of all records was conducted and no failures during production were reported.A copy of the test report will be filed with the fda to close the mdr report filed for this issue.Without an examination of the suspect device, the factory was unable to determine a root cause, and no corrective action will be implemented at this time.No similar complaints have been received within the past 12 months reviewed.We will continue to monitor our feedback system and take any additional action as necessary.We consider this report closed.
 
Event or Problem Description
It was reported by the physician that the needle disconnected from the hub during a pelvis injection and was left in the patient.It was also reported this occured three (3) times and he has used these needles previously.He did state the area in the pelvis required him to push hard into the patient because it was a narrow injection area and when he pulled the needle out it disconnected.All three patients were reported to be moderately obese.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELI
Common Device Name
ANESTHESIA NEEDLE
Manufacturer (Section D)
DR. JAPAN CO., LTD.
4-15-1 nagano
gyoda city, saitama 361-0
JA  361-0
MDR Report Key10950317
Report Number1058382-2020-00003
Device Sequence Number19887220
Product Code BSP
UDI-Device Identifier00386120006758
UDI-Public00386120006758
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/04/2020
Report Date (Section F) 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/21/2024
Device Model NumberSN22G351
Device Catalogue NumberSN22G351
Device Lot Number1903758
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date11/04/2020
Device Age12 MO
Event Location Other
Date Report to Manufacturer11/04/2020
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1