| Model Number |
PMXENGN-A |
| Medical Device Problem Code |
Decrease in Suction (1146)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2020-02181.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a thrombectomy procedure for a pulmonary embolism using a penumbra engine (engine) and penumbra engine canister (canister).During the procedure, the physician turned on the engine and noticed the aspiration power to be low.The physician removed the canister and replaced it with a new canister to troubleshoot; however, it was unsuccessful.Therefore, the engine and canister was removed.The procedure was completed using another engine.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Please note that the following sections are being updated based on additional information received from the device return on 01/11/2021: 1.Section d.Box 4.Serial #; 2.Section h.Box 4.Device manufacture date.Evaluation of the returned engine revealed a functional device.The engine was able to power on and produced vacuum pressure within specification.The engine remained powered on for approximately 5 minutes and no decrease in aspiration was observed.The reported complaint could not be confirmed.Penumbra engines are inspected at incoming quality control which includes a visual inspection as well as a verification of test results to ensure specifications for each output are met.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|
| |
|
Search Alerts/Recalls
|