• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. ULTRASITE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. ULTRASITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 470021
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as event 2 of b.Braun medical internal report number (b)(4).The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
Event 2: blood leakage was reported.As reported by the user facility: "leakage occurred at the connection between the ext set male spin lock connector and the iv site.".
 
Additional Manufacturer Narrative
Event 2: this report has been identified as b.Braun medical internal report number (b)(4).One (1) video was provided for evaluation.Visual examination of the video noted leakage occurring between the extension set and catheter female/male connection.It is unable to be determined whether the extension set or catheter contributed to the defect observed.Although the reported defect of leakage was confirmed without the sample the exact root cause cannot be determined at this time.  review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.  we will maintain this report for further references and continue to monitor other reports for similar occurrences.  if any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd
allentown 18109
MDR Report Key10950798
Report Number2523676-2020-00339
Device Sequence Number2882088
Product Code FPA
UDI-Device Identifier04046964183850
UDI-Public04046964183850
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K955585
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model Number470021
Device Catalogue Number470021
Device Lot Number0061746475
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/20/2020
Supplement Date Received by Manufacturer11/20/2020
Initial Report FDA Received Date12/05/2020
Supplement Report FDA Received Date01/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-