The reported event could be confirmed as the plates (4 fragments) were returned.The postoperative plate breakages could be observed at the fixing holes of the implants.No further physical inspections were performed as the reported event was assessed by stryker¿s medical expert as off-label use.A follow-up with the sales rep has confirmed that the surgeon has used the reported plates for condylar process fractures.The devices in question are part of the mid-face fixation module and are not intended to be used for mandible fixation.The mechanical strength of the reported plates is comparably lower to that of the plates intended for mandible fixation.As shown in the related brochure (universal 2, 9410-400-184 rev.None), the minimum profile height of a plate for mandible fixation is 1.0mm, whereas the plate 55-06704 has a profile height of 0.6mm.The difference of the profile heights between the mid-face plate and mandible plate may have influenced the determined plate breakage.Stryker¿s medical expert stated that ¿this is off-label use.The mid face plates are for non-load bearing fractures; the mandible is load bearing.¿in sum, the procedure to fixate mandible fractures with plates intended for mid-face fixation is considered as off-label use.No lot numbers were provided and a review of the specific manufacturing batch record and related quality documents (manufacturing documents, inspection plan, inspection drawing and release report) cannot be performed.Therefore, it is also not possible to determine the quantities released for distribution within the affected lots.Based on the investigation, there is no indication for an incorrectly working product or any systematic design, material or manufacturing related issue.Therefore, no further corrective and / or preventive actions are deemed necessary at this time.The complaint is added to the complaint trend.
|