Additional information was added to d10, h3, h4 and h6.H10: the actual device was not available; however, six (6) units were received for evaluation.During visual inspection, damage was observed on the interlink t-conn, non-retractable male luer lock.Functional testing was performed including pressure and clear passage testing.During clear passage testing, the device performed according to product specifications.During pressure testing, the device failed at interlink t-conn, non-retractable male luer lock in all six devices.The reported condition was verified.The cause of the condition was manufacturing related.This issue is being further investigated.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|