Model Number TRL-510
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Visual analysis of the returned device identified broken shell.Weight measurement identified device was underweight.A microscopic analysis was performed which identified sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identified the thickness within specification.Based on the device analysis, the final assessment is a sharp broken on posterior due to an unidentified (tear) opening and a sharp broken on anterior due to an unidentified (tear) opening.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Patient reported right side "lump in breast" clarified as scar tissue and not related to the device.Healthcare professional reported rupture, ¿bottoming out of the breast", ¿soreness", "discomfort¿, ¿lump in breast", ¿feels too heavy and feels really heavy laterally¿ and "recall implant".Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key10951585
Report Number9617229-2020-21592
Device Sequence Number3368888
Product Code FTR
UDI-Device Identifier10888628007772
UDI-Public10888628007772
Combination Product (Y/N)N
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/02/2021
Device Model NumberTRL-510
Device Catalogue NumberTRL-510
Device Lot Number2905781
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2019
Initial Report FDA Received Date12/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-