| Model Number |
9-PDA-007 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/28/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, a 12/10mm amplatzer duct occluder was selected for implant.During the procedure, both the discs "did not open properly" and "failed with repeated attempt." therefore, the device was removed from the patient and a new 12/10mm amplatzer duct occluder was successfully implanted.No patient consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of both discs not opening properly could not be confirmed.The investigation confirmed the device met visual and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
|
| |
|