| Model Number |
GB109 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
| Date of Event |
11/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported to aesculap ag that a microline straight handpiece (part # gb109) was used during an oral surgery procedure performed on (b)(6) 2020.According to the complainant, the device generated heat when used in combination with another manufacturers bar.Reportedly, a large swollen burn occurred on the back of the patient's lips.The complaint device was available to be returned to the manufacturer for evaluation.Additional information was requested, but has not been made available.The adverse event / malfunction is filed under aag reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to the manufacturer for physical evaluation.A visual examination of the returned device revealed, the handpiece to be in very bad condition.The interior was very dirty and the front ball bearing was broken.Overheating would be likely, due to the defective ball bearing.Which would have resulted in an inability to clamp the tool properly.The investigation into the cause of the reported problem was able to confirm the failure mode of overheating.The complaint investigator identified a damaged front ball bearing.Which likely resulted, in the handpiece overheating.As such, the reported complaint of device overheating was confirmed.
|
| |
|