| Model Number |
BXA095902A |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The engineering evaluation is currently in progress.
|
| |
|
Event or Problem Description
|
|
The following was reported by the gore fsa: the patient presented for a snorkel case.The gore® viabahn® vbx balloon expandable endoprosthesis was advanced into the celiac artery.The physician was only able to obtain 4 atmospheres during the inflation of the device.The physician was able to remove the balloon and insert a different balloon, successfully deploying the gore® viabahn® vbx balloon expandable endoprosthesis.There was no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Conclusion code updated.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturer narrative: cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.H.6.Results code 1: 213: a review of the manufacturing records verified the lot met all pre-release specifications.H.6.Results code 2: 213 the engineering analysis stated the following: the as-reported complaint, partial deployment, cannot be directly confirmed because the endoprosthesis remains implanted.Balloon leakage is confirmed and is considered the cause of reported partial deployment.The cause of the damage to the balloon cannot be further identified.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2020 a patient presented for a snorkel case.A gore® viabahn® vbx balloon expandable endoprosthesis was advanced into the celiac artery.During the inflation of the device, only 4 atmospheres could be obtained.The vbx balloon was removed and replaced with a different balloon, successfully deploying the gore® viabahn® vbx balloon expandable endoprosthesis.There was no harm to the patient.
|
| |
|