| Model Number |
53612 |
| Medical Device Problem Code |
Improper Chemical Reaction (2952)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported three of the catheters were not as "smooth" or "coated" and they were difficult to insert, the lubricant was thicker.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed.There was no definitive root cause determined for the increase in lubricity complaints.It was determined however, that there were several confounding causes that could contribute to lubricity complaints.The fishbone diagram identifies those contributing factors to lubricity and increased complaints rules out categories from being a root cause of the failure mode.The device history record review was not required as the reported event was confirmed capa case.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported three of the catheters were not as ¿smooth¿ or ¿coated¿and they were difficult to insert.It was also stated that the lubricant was thicker.
|
| |
|
Search Alerts/Recalls
|