• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. MICROCUFF SUBGLOTTIC SUCTIONING ENDOTRACHEAL TUBE, 8.0 MM, ORAL; VAP MICROCUFF ENDOTRACHEAL TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. MICROCUFF SUBGLOTTIC SUCTIONING ENDOTRACHEAL TUBE, 8.0 MM, ORAL; VAP MICROCUFF ENDOTRACHEAL TUBES Back to Search Results
Model Number 13222
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 04-dec-2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
It was reported there was and incident of the cuff not holding air.The cuff was not inflating on intubation.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 20031697, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.The device was evaluated and the failure was confirmed.Root cause could not be determined.All information reasonably known as of 23-dec-2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROCUFF SUBGLOTTIC SUCTIONING ENDOTRACHEAL TUBE, 8.0 MM, ORAL
Common Device Name
VAP MICROCUFF ENDOTRACHEAL TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key10954143
Report Number9611594-2020-00241
Device Sequence Number16946551
Product Code BTR
UDI-Device Identifier00609038132228
UDI-Public00609038132228
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K131254
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/03/2022
Device Model Number13222
Device Catalogue NumberN/A
Device Lot Number20031697
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/10/2020
Supplement Date Received by Manufacturer12/09/2020
Initial Report FDA Received Date12/05/2020
Supplement Report FDA Received Date12/23/2020
Patient Sequence Number1
-
-