H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one dorado pta catheter was returned for evaluation.On the visual evaluation, the device have some blood residue.The balloon was still in folded position within balloon protector and balloon noted to be bunched.Circumferential break noted at the glue joint and the inner inflation lumen was exposed.No other anomalies were noted to the device.Therefore, the investigation has confirmed for material deformation balloon noted to be bunched distally on the device returned for evaluation.The investigation has confirmed for identified break as the returned device noted to have break at the proximal glue joint and the inner guide wire lumen was exposed.A definitive root cause for the material def.Ormation and identified break could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4(expiry date: 05/2021),g3,h6(method).H11: h6(device, result and conclusion).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|