MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL OG TDL CMPLX FILL COIL 5X15; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL OG TDL CMPLX FILL COIL 5X15; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 640CF0515
Medical Device Problem Codes Failure to Advance (2524); Difficult to Advance (2920); Positioning Problem (3009); Premature Separation (4045)
Health Effect - Clinical Codes Obstruction/Occlusion (2422); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.Initial reporter: the customer contact information, including name, occupation, phone, fax, and e-mail address, was not reported.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.This is one of three products involved with the complaint and the associated manufacturer report numbers are 3008114965-2020-00563, and 3008114965-2020-00564.Complaint conclusion:a report from the field indicated that during a coil embolization of an intracranial aneurysm, a 5mm x 15cm orbit galaxy tdl complex fill (640cf0515/30298966) coil prematurely detached and released to the middle and anterior cerebral artery.A sab640 stent retriever was used in attempt to retrieve the coil but was unsuccessful.Therefore, a 4.5mm x 22 enterprise no tip enterprise (enc452200/11213047) stent was implanted via a prowler select plus (606-s255x/30378173) microcatheter at the middle cerebral artery (mca) to secure the coil to the vessel wall.It was reported that the physician encountered resistance while advancing the coil through the headway 17 (microvention) microcatheter and the 6f envoy (67025800/30232241) guide catheter subsequently moved from the c2 to the m1.It was further stated that this may have caused the coil to unintendedly release.The patient has been awake.The vessels were considered normal, without acute bends or excessive tortuosity.Additional information received indicated that post-procedure angiography showed smooth blood flow in the right anterior cerebral artery (aca) and the original aneurysm was no longer developed.Thrombosis was noted within the aneurysm.There was a right posterior cerebral artery occlusion.The patients right upper limb muscle strength was level 0 and right lower limb muscle strength was level 1.Vital signs were stable.The patient was transferred to another hospital on the following day.A neuroscout (601-314/j6867r) guidewire and trufill ii (635-002/96331364) syringe were also used during the procedure.Imaging/films were provided.Additional event information received on 20-nov-2020 indicated that the initial aneurysm embolization was performed on (b)(6) 2020, and the complaint coil was left in the patient.The patient was transferred to another hospital on (b)(6) 2020, and underwent a new procedure on (b)(6) 2020 to remove the stent and coil.The complaint coil and stent were successfully removed with a stent retriever.The target aneurysm was unruptured, however, its exact anatomical location and characteristics were not reported.It is unknown if neck remodeling was utilized.The complaint coil prematurely detached near the distal tip of the concomitant microcatheter during advancement.It is not known if the operator was able to move the coil at all due to the resistance or if the device was stuck.The issue required removal of the coil delivery tube and microcatheter as a unit, resulting in a loss of cerebral target position.The coil did not become unraveled or stretched prior to the unintended detachment.There was an adequate and continuous flush maintained through the devices.There had been no attempts to detach the coil prior to the event.The concomitant devices functioned as intended.However, it was reported that resistance was encountered during use of the enterprise stent (pending clarification).The stent fully expanded and there was good apposition between all areas of the stent and the vessel.It is not known whether the event resulted in restriction of blood flow, or thrombosis; however, the event did lead to prolongation of existing hospitalization.It was last reported that the patient is in stable condition.Additional information received on 23-nov-2020 indicated that the target aneurysm was located at the anterior cerebral artery (aca).Resistance was a reported to be a device performance issue associated with the enterprise stent.The resistance was encountered during the initial procedure on(b)(6) 2020.It was unknown what was happening procedurally when the resistance occurred, and it was unknown if the resistance was felt during advancement of the enterprise vrd through the prowler select plus microcatheter.This issue was resolved as the stent and the coil of accidental release were successfully removed with the thrombectomy stent on (b)(6) 2020.The patient tolerated the removal well.This issue required removal of the enterprise vrd and prowler select plus mc, resulting in a loss of cerebral target position.The enterprise stent and the prowler select plus mc were both not replaced.The patient was discharged however the disposition of the discharge is unknown.No further information was provided.The device was not returned for analysis, therefore, no further investigation can be performed.Mre review: a manufacturing record evaluation was performed for the finished device 30298966 number, and no non-conformances related to the reported complaint condition were identified.Returned imaging/films were reviewed by independent physician and the findings are as follows: it is unclear what they are intending to treat because there is no imaging provided of the aneurysm.The coils seen on the imaging are loosely configured, rather than a tight coil mass that is typically seen in the treatment of an aneurysm.Given the information above, it has been assumed this is because it is the first coil that was attempted, and it did not seat into the aneurysm.A determination cannot be made regarding the coil selection because there is no information provided about the aneurysm size.The imaging suggests that the coil is likely in the aca/mca and perhaps in the distal ica, however the angle of the image acquisition makes it challenging to confirm.There is also an image of the microcatheter in the cervical region within the guide catheter and the coil is distal to the tip of the microcatheter.There is also a video supplied, but it is unclear what it is attempting to depict.Based on the information provided, it is difficult to determine a causal relationship.Pre-mature coil detachment can occur with the coil during priming or during deployment/re-positioning.There is not enough information to infer what occurred.Resistance/friction during advancement and unintended coil detachment are known potential procedural complications associated with the orbit galaxy coil.The orbit galaxy instructions for use (ifu) cautions the operator to never advance, withdraw, or torque the delivery tube against resistance without first determining the cause of resistance under fluoroscopy.Manipulation of the delivery tube against resistance can cause damage and/or premature detachment of the coil.If unusual friction is noted within the infusion catheter, remove the detachable coil system.If friction is noted with any subsequent detachable coil system, carefully examine the detachable coil system and the infusion catheter for possible damage.Replace both if necessary.Excessive force applied to a coil through repeated coil manipulation can increase the possibility of premature detachment.Review of the limited information does not allow for an exact determination of root cause.However, there are clinical and procedural factors, including aneurysm/vessel characteristics, device selection, device interaction, and operator technique, that may have contributed to the reported event.The file will be re-reviewed if additional information is received at a later date.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
 
Event or Problem Description
A report from the field indicated that during a coil embolization of an intracranial aneurysm, a 5mm x 15cm orbit galaxy tdl complex fill (640cf0515/30298966) coil prematurely detached and released to the middle and anterior cerebral artery.A sab640 stent retriever was used in attempt to retrieve the coil but was unsuccessful.Therefore, a 4.5mm x 22 enterprise no tip enterprise (enc452200/11213047) stent was implanted via a prowler select plus (606-s255x/30378173) microcatheter at the middle cerebral artery (mca) to secure the coil to the vessel wall.It was reported that the physician encountered resistance while advancing the coil through the headway 17 (microvention) microcatheter and the 6f envoy (67025800/30232241) guide catheter subsequently moved from the c2 to the m1.It was further stated that this may have caused the coil to unintendedly release.The patient has been awake.The vessels were considered normal, without acute bends or excessive tortuosity.Additional information received indicated that post-procedure angiography showed smooth blood flow in the right anterior cerebral artery (aca) and the original aneurysm was no longer developed.Thrombosis was noted within the aneurysm.There was a right posterior cerebral artery occlusion.The patients right upper limb muscle strength was level 0 and right lower limb muscle strength was level 1.Vital signs were stable.The patient was transferred to another hospital on the following day.A neuroscout (601-314/j6867r) guidewire and trufill ii (635-002/96331364) syringe were also used during the procedure.Imaging/films were provided.Additional event information received on 20-nov-2020 indicated that the initial aneurysm embolization was performed on (b)(6) 2020, and the complaint coil was left in the patient.The patient was transferred to another hospital on (b)(6) 2020, and underwent a new procedure on (b)(6) 2020 to remove the stent and coil.The complaint coil and stent were successfully removed with a stent retriever.The target aneurysm was unruptured, however, its exact anatomical location and characteristics were not reported.It is unknown if neck remodeling was utilized.The complaint coil prematurely detached near the distal tip of the concomitant microcatheter during advancement.It is not known if the operator was able to move the coil at all due to the resistance, or if the device was stuck.The issue required removal of the coil delivery tube and microcatheter as a unit, resulting in a loss of cerebral target position.The coil did not become unraveled or stretched prior to the unintended detachment.There was an adequate and continuous flush maintained through the devices.There had been no attempts to detach the coil prior to the event.The concomitant devices functioned as intended.However, it was reported that resistance was encountered during use of the enterprise stent (pending clarification).The stent fully expanded and there was good apposition between all areas of the stent and the vessel.It is not known whether the event resulted in restriction of blood flow or thrombosis, however, the event did lead to prolongation of existing hospitalization.It was last reported that the patient is in stable condition.Additional information received on (b)(6) 2020, it indicated that the target aneurysm was located at the anterior cerebral artery (aca).Resistance was a reported to be a device performance issue associated with the enterprise stent.The resistance was encountered during the initial procedure on (b)(6) 2020.It was unknown what was happening procedurally when the resistance occurred, and it was unknown if the resistance was felt during advancement of the enterprise vrd through the prowler select plus microcatheter.This issue was resolved as the stent and the coil of accidental release were successfully removed with the thrombectomy stent on (b)(6) 2020.The patient tolerated the removal well.This issue required removal of the enterprise vrd and prowler select plus mc, resulting in a loss of cerebral target position.The enterprise stent and the prowler select plus mc were both not replaced.The patient was discharged, however, the disposition of the discharge is unknown.No further information was provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Updated sections on this medwatch report: b5, g3, g6, h2, h6, h10 and concomitant products.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Section b5: revised event narrative was provided by the local bq team on (b)(6) 2021.The updated event description is as follows: during a coil embolization of an unruptured left anterior communicating artery aneurysm, a ,5mm x 15cm orbit galaxy tdl complex fill (640cf0515/30298966) coil exhibited resistance at the time of aneurysm basket formation.The physician continued to feed the coil into the aneurysm sac, but the coil could not proceed.The concomitant headway 17 (microvention) microcatheter was withdrawn, but the coil prematurely detached.The coil immediately caused an occlusion at the left anterior communicating artery and left side of the brain.The physician attempted to capture the prematurely detached coil with a solitaire 630 (medtronic) stent retriever, but this was unsuccessful.The physician considered implanting an enterprise stent to improve the blood flow in the left mca, but ultimately decided to implant the solitaire stent retriever distal to the left mca and proximal to the left ica.Angiography revealed that the blood flow of the left mca had been restored, and the left aca was occluded.It was further reported the embolization of the target aneurysm was not able to be completed.A 6f envoy guiding catheter was used in conjunction with the complaint coil.Revised complaint conclusion: during a coil embolization of an unruptured left anterior communicating artery aneurysm, a ,5mm x 15cm orbit galaxy tdl complex fill (640cf0515/30298966) coil exhibited resistance at the time of aneurysm basket formation.The physician continued to feed the coil into the aneurysm sac, but the coil could not proceed.The concomitant headway 17 (microvention) microcatheter was withdrawn, but the coil prematurely detached.The coil immediately caused an occlusion at the left anterior communicating artery and left side of the brain.The physician attempted to capture the prematurely detached coil with a solitaire 630 (medtronic) stent retriever, but this was unsuccessful.The physician considered implanting an enterprise stent to improve the blood flow in the left mca, but ultimately decided to implant the solitaire stent retriever distal to the left mca and proximal to the left ica.Angiography revealed that the blood flow of the left mca had been restored, and the left aca was occluded.It was further reported the embolization of the target aneurysm was not able to be completed.A 6f envoy guiding catheter was used in conjunction with the complaint coil.The device was not returned for analysis; therefore, no further investigation can be performed.Mre review: a manufacturing record evaluation was performed for the finished device 30298966 number, and no non-conformances related to the reported complaint condition were identified.Returned imaging/films were reviewed by independent physician and the findings are as follows: it is unclear what they are intending to treat because there is no imaging provided of the aneurysm.The coils seen on the imaging are loosely configured, rather than a tight coil mass that is typically seen in the treatment of an aneurysm.Given the information above, it has been assumed this is because it is the first coil that was attempted, and it did not seat into the aneurysm.A determination cannot be made regarding the coil selection because there is no information provided about the aneurysm size.The imaging suggests that the coil is likely in the aca/mca and perhaps in the distal ica, however the angle of the image acquisition makes it challenging to confirm.There is also an image of the microcatheter in the cervical region within the guide catheter and the coil is distal to the tip of the microcatheter.There is also a video supplied but it is unclear what it is attempting to depict.Based on the information provided, it is difficult to determine a causal relationship.Pre-mature coil detachment can occur with the coil during priming or during deployment/re-positioning.There is not enough information to infer what occurred.Poor coil conformability, impeded in microcatheter, and unintended coil detachment are known potential procedural complications associated with the orbit galaxy coil.The orbit galaxy instructions for use (ifu) cautions the operator to never advance, withdraw, or torque the delivery tube against resistance without first determining the cause of resistance under fluoroscopy.Manipulation of the delivery tube against resistance can cause damage and/or premature detachment of the coil.If unusual friction is noted within the infusion catheter, remove the detachable coil system.If friction is noted with any subsequent detachable coil system, carefully examine the detachable coil system and the infusion catheter for possible damage.Replace both if necessary.Excessive force applied to a coil through repeated coil deployments or manipulation can increase the possibility of premature detachment.Review of the available information does not allow for an exact determination of root cause.However, there are clinical and procedural factors, including aneurysm/vessel characteristics, device selection, device interaction, and operator technique that may have contributed to the reported event rather than the design or manufacture of the device.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OG TDL CMPLX FILL COIL 5X15
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
MDR Report Key10958156
Report Number3008114965-2020-00562
Device Sequence Number16986430
Product Code KRD
UDI-Device Identifier10886704030300
UDI-Public10886704030300
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K093973
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Model Number640CF0515
Device Catalogue Number640CF0515
Device Lot Number30298966
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2020
Supplement Date Received by Manufacturer06/30/2021
Initial Report FDA Received Date12/07/2020
Supplement Report FDA Received Date07/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENVOY 6F, MPD 100CM; EU ENT4.5MMD 22MML WNO DSTL TP; HEADWAY 17 MICROCATHETER; NEUROSCOUT 14, STD; PROWLER SELECT PLUS 150/5CM; SOLITAIRE 630 (MEDTRONIC) STENT RETRIEVER; TRUFILL DCS SYRINGE II
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-