| Model Number |
WA90004W |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bowel Perforation (2668)
|
| Date of Event |
12/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the patient's outcome and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
|
| |
|
Event or Problem Description
|
|
Olympus was informed that during an unspecified procedure the patient sustained a minor perforation of the colon wall.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect medical device was not returned to the manufacturer for evaluation/investigation but was examined by an olympus service technician at the customer facility without finding any malfunctions of the argon plasma coagulation unit.There were also no reports of any malfunctions of the argon plasma coagulation unit or any of the concomitant medical devices.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected serial number of the argon plasma coagulation unit without showing any abnormalities.Based on the information available, the exact cause of the reported phenomenon and the patient¿s outcome could not be determined and is being judged as unknown.However, there is no indication that a technical cause led to the reported patient injury, rather it can most likely be attributed to the patient coughing during the procedure as described in the event description.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
|
| |
|
Search Alerts/Recalls
|