• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND SELECTSITE; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND SELECTSITE; CATHETER, PERCUTANEOUS Back to Search Results
Model Number C304-HIS
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during an implant procedure it took a lot of force to slit the catheter.It was additionally noted that the wire came out of the sheath and was exposed.The catheter was used to complete the implant.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the delivery catheter was returned and analyzed.Analysis indicated the mechanical operation of the catheter slitting showed a spiral slit.The catheter did not peel along the score lines.The manipulator wire pulled through the lumen wall.The mechanical operation of the catheter shaft was kinked/buckled.Visual analysis of the catheter indicated damage during use.The analyst noted it is likely that the slitting was not on score line causing spiral slit.Spiral slitting led to the catheter deflectable wire being pull out of the lumen wall.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELECTSITE
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
MDR Report Key10959772
Report Number9612164-2020-04795
Device Sequence Number3489139
Product Code DQY
UDI-Device Identifier00763000147006
UDI-Public00763000147006
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K183493
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/17/2022
Device Model NumberC304-HIS
Device Catalogue NumberC304-HIS
Device Lot Number0010375516
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2021
Initial Date Received by Manufacturer 11/23/2020
Supplement Date Received by Manufacturer02/17/2021
Initial Report FDA Received Date12/07/2020
Supplement Report FDA Received Date02/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age81 YR
Patient Weight100
-
-