• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number 2AF284
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a cryoablation procedure, bubbles were noted in the balloon catheter when pulling it back.The balloon catheter was replaced, resolving the issue.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Correction: d3 product event summary: the balloon catheter 2af284 with lot number 00929 was returned and analyzed.Visual inspection showed that the catheter was intact with no apparent issues.Smart chip verification indicates that the catheter has not been used.Without reprogramming the catheter, it was connected to the console and no system notices were received; catheter was recognized.The catheter passed performance test and the electrical continuity test.Pin-to-pin impedance measurement did not show any fluctuation.In conclusion, the reported air ingress was not confirmed through testing.The balloon catheter passed the returned product inspection.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Common Device Name
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key10959972
Report Number3002648230-2020-00636
Device Sequence Number16941285
Product Code OAE
UDI-Device Identifier00643169878365
UDI-Public00643169878365
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P100010/S015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/10/2021
Device Model Number2AF284
Device Catalogue Number2AF284
Device Lot Number00929
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2021
Initial Date Received by Manufacturer 11/30/2020
Supplement Date Received by Manufacturer03/01/2021
Initial Report FDA Received Date12/07/2020
Supplement Report FDA Received Date03/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-