Model Number 9438-05
Medical Device Problem Code No Device Output (1435)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that transmitter failed error occurred.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that transmitter failed error occurred.The product was evaluated.An external visual inspection was performed and passed.Voltage test was performed and failed due to 0v.No data was provided for evaluation. confirmation of the allegation was undetermined.The probable cause could not be determined. no injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
MDR Report Key10960173
Report Number3004753838-2020-150293
Device Sequence Number3489762
Product Code MDS
UDI-Device Identifier00386270000125
UDI-Public00386270000125
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/27/2020
Device Model Number9438-05
Device Catalogue NumberSTT-GL-004
Device Lot Number5271888
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/13/2020
Supplement Date Received by Manufacturer03/24/2021
Initial Report FDA Received Date12/07/2020
Supplement Report FDA Received Date03/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-