| Model Number |
3507800BC |
| Medical Device Problem Codes |
Material Rupture (1546); Migration (4003)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Fall (1848)
|
| Date of Event |
11/20/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture and rupture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old caucasian female patient underwent a breast augmentation primary with a mentor memorygel breast implant 800cc and suffered from capsular contracture and rupture symptomatic on the left breast implant.As a result, the patient has undergone removal and replacement with non-mentor device on (b)(6) 2020.
|
| |
|
Additional Manufacturer Narrative
|
|
On 12/9/2020, it was reported to mentor that capsular contracture was baker grade iv.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 12/17/2020, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 1/6/2021, during the visual inspection, the sample was found to be ruptured and received in two parts.Additionally, an anomaly was observed on the edges of the tear consistent with a crease/fold.The evaluation determined that the loss of shell integrity is consistent with a crease fold rupture of the implant shell, which is a known inherent risk of gel-filled mammary prosthesis.Rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in a rupture in a later time.Breast implants may also simply wear out over time.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination.As part of our quality process, the manufacturing records of this lot-serial number were reviewed and the manufacturing standards were met prior to the release of this lot.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Complaint information is consistently analyzed and monitored by mentor quality assurance to determine when further action is necessary.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 05/10/2021, it was reported to mentor that the patient experienced baker grade iv capsular contracture on the right side breast.Possible left side rupture secondary to trauma ¿ chest injury ((b)(6) 2020 ¿ severe motorcycle accident), now with malposition and capsular contracture.The replacement devices were allergan brand scx gel implants.The patient had undergone removal of left ruptured silicone implant with some extracapsular silicone extrusion, reconstruction and repair of left pectoralis major muscle on (b)(6) 2020.Manufacturer¿s reference number: (b)(4).
|
| |
|