Catalog Number 106697
Device Problem Obstruction of Flow (2423)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/11/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that after six hours of continuous renal replacement therapy using a prismaflex m100 with citrate anticoagulant, the set clotted."lipid or fibrinoid kind of deposition" was observed on the filter top and other parts of the deaeration chamber and return line.Treatment was restarted with a new m100 set using heparin as anticoagulant.Within an hour, filter pressure reached above 200 and tmp above 100 and early clotting of the filter was observed.No additional information is available.
 
Manufacturer Narrative
Additional information: h3, h6 and h10.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection of the provided picture could not identify any specific defect on the impacted set.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M100
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10960901
MDR Text Key224635939
Report Number8010182-2020-00284
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
PMA/PMN Number
K041005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number106697
Device Lot Number19E0401Z
Was Device Available for Evaluation? No
Date Manufacturer Received01/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
BIPHOZYL CRRT
Patient Age69 YR
Patient Weight80
-
-