• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX M100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX M100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955595
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 10/14/2020
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone no.: (b)(6).Facility name: (b)(6) hospital.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
I was reported that during priming with a prismaflex m100 ,an external fluid leakage was observed.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information: h6 and h10.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection of the provided picture showed that there was an assembly issue with the return line post deaeration chamber.The reported condition was verified.The cause was due to an observed misassembly.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M100
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10960939
MDR Text Key225548432
Report Number8010182-2020-00285
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414124397
UDI-Public(01)07332414124397
Combination Product (y/n)Y
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/01/2022
Device Catalogue Number955595
Device Lot Number20B1904
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/12/2020
Initial Date FDA Received12/07/2020
Supplement Dates Manufacturer Received12/28/2020
Supplement Dates FDA Received01/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-