Device was received and evaluated.Evaluation determined that the device was found with cut on the ¿a-rubber¿ and a loose bcu (body control unit) was observed.Broken bending section was determined and intermittent image was observed.In addition, peeling on ¿a-rubber¿ glue was determined and switches are found not working.The components of the device were repaied and restored to olympus original factory specifications.Based on evaluation findings the reported failure was confirmed and failures found likely attributed to users handling and or maintenance issue.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|