• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1070-080
Medical Device Problem Codes Material Rupture (1546); Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion located in an unspecified vessel that was heavily calcified.After a bare metal stent was deployed, the lesion was post dilated with an armada 35 balloon dilatation catheter which ruptured circumferentially during the first inflation below rated burst pressure.There was no reported adverse patient effect and no clinically significant delay in the procedure.Another armada balloon was used to complete the procedure.No additional information was provided.
 
Event or Problem Description
Subsequent to the previously filed reports, a cutdown procedure was performed to remove all device material from the patient.No additional information was provided.
 
Additional Manufacturer Narrative
Visual analysis was performed on the returned device.The reported balloon rupture and separation were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined that the reported difficulties are due to case circumstances.It is likely that the balloon rupture and material separation were caused by the interaction of the bare metal stent and heavy calcification.The reported treatments appear to be related to the operational context of the procedure as the device material was surgically removed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10963471
Report Number2024168-2020-10261
Device Sequence Number11343526
Product Code LIT
UDI-Device Identifier08717648154805
UDI-Public08717648154805
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2021
Device Model NumberB1070-080
Device Catalogue NumberB1070-080
Device Lot Number80109G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2020
Initial Date Received by Manufacturer 11/16/2020
Supplement Date Received by Manufacturer12/15/2020
Initial Report FDA Received Date12/07/2020
Supplement Report FDA Received Date12/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient Weight90
-
-