| Model Number |
AC*T DIFF 2 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) assisted the customer via telephone.Controls were passing, it seems that they changed qc lots but didn't adjust their ranges for approximately two months.The customer was not a med tech and did not understand the assayed control issue.The fse verified the instrument function with the customer by running fresh tubes of 4c-es controls an reviewed the data with the customer.Recommended to the customer that they should calibrate the instrument whenever they encounter this issue and after any manual calibration.Non instrument malfunction could be identified by the fse.Bec internal identifier - case-(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported erroneous high platelet (plt) results on patient samples run on the act diff 2 analyzer after onsite service.There were no erroneous patient results reported outside the laboratory.There was no report of impact to patient treatment.
|
| |
|