| Model Number |
227-5 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Pulmonary Dysfunction (2019)
|
| Date of Event |
11/11/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A piece of the sample was returned and is being process for evaluation.The device history record for lot m18170t302 was reviewed and the product was produced according to product specifications.All information reasonably known as of 02 dec 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's state of health began to deteriorate.An acute bronchoscopy was performed, and a foreign body was removed from the lungs.The foreign body is alleged to be from the valve flap of the closed suction catheter system.Additional information has been requested but not yet received.
|
| |
|
Additional Manufacturer Narrative
|
|
The sample returned consisted only of the flapper valve material that was detached from the assembly.The rest of the closed suction catheter was not returned.The material of the flapper valve was wrinkled in appearance with some cracking on the component.Attempts were made to reproduce this effect during manufacturing; however, the condition of the returned sample was unable to be duplicated.The complaint was confirmed as reported, however, no root cause could be determined.All information reasonably known as of 30 mar 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Search Alerts/Recalls
|