The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to determine if any product condition could have contributed to the skin irritation.Lot release records were reviewed and the product lot met all acceptance criteria.The omnipod is iso10993 compliant, undergoing cytotoxicity, sensitization, genotoxicity, hemolysis, irritation or intracutaneously reactivity, systemic toxicity, and implantation effects testing.The omnipod is sterilized with 100% ethylene oxide gas with a sterility assurance level of 10-6 per iso11135 and eo residual levels in compliance with iso10993.Each lot is confirmed to meet requirements for non-pyrogenic per iso10993 and sterility per iso11135 prior to release.
|
It was reported that after wearing the pod on the leg between 36 and 48 hours, the site was bleeding, inflamed infected with pus, with high blood glucose (bg) levels of 480 mg/dl.The patient visited the emergency room (er) where was prescribed a cream (ichtholan) to be used on the affected area.A new pod was applied to treat the hyperglycemia.
|