Catalog Number 532.010
Medical Device Problem Code Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.Udi: (b)(4).
 
Event or Problem Description
It was reported that during an unspecified surgical procedure, it was observed that the small battery drive device was either smoking or would not run.The reporter was unsure about which one occurred.There were no reports of surgical delay.It was not reported if a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in (b)(6) 2020.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device safety disk was missing, the unit would not run and components were missing.The device also failed pretests for check the off/oscillation/on switch mode function and check the function with electric pen drive console.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component failure due to normal wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key10968664
Report Number8030965-2020-09441
Device Sequence Number17213943
Product Code HWE
UDI-Device Identifier07611819110779
UDI-Public07611819110779
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K971544
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number532.010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2020
Supplement Date Received by Manufacturer01/11/2021
Initial Report FDA Received Date12/08/2020
Supplement Report FDA Received Date01/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No