• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7015
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at the time of event: 18 years or older.
 
Event or Problem Description
It was reported that shaft break occurred.The 98% stenosed, 20mmx2.75mm target lesion was located in the severely tortuous and severely calcified right cronary artery.A 3.0mm x 15mm quantum maverick balloon was advanced for dilatation.However, the shaft was kinked and upon removal, the shaft was fractured outside the patient's body.The procedure was completed with a different device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
Age at the time of event: 18 years or older.Device evaluated by mfr.: the device was returned for analysis.Returned product consisted of a quantum maverick mr balloon catheter, in two pieces.The balloon was tightly folded.Analysis of the tip, balloon, inner/outer shaft and hypotube included microscopic and visual inspection.Inspection revealed a complete separation in the hypotube that was located 68.2cm from the strain relief, and there were numerous kinks in the hypotube.Inspection of the rest of the device found no other damage or defects.
 
Event or Problem Description
It was reported that shaft break occurred.The 98% stenosed, 20mmx2.75mm target lesion was located in the severely tortuous and severely calcified right cronary artery.A 3.0mm x 15mm quantum maverick balloon was advanced for dilatation.However, the shaft was kinked and upon removal, the shaft was fractured outside the patient's body.The procedure was completed with a different device.There were no patient complications reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10969209
Report Number2134265-2020-17168
Device Sequence Number11815753
Product Code LOX
UDI-Device Identifier08714729392781
UDI-Public08714729392781
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2023
Device Model Number7015
Device Catalogue Number7015
Device Lot Number0024978952
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2020
Initial Date Received by Manufacturer 11/18/2020
Supplement Date Received by Manufacturer01/14/2021
Initial Report FDA Received Date12/08/2020
Supplement Report FDA Received Date02/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-