| Model Number |
07P7430 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated ldh results generated on the alinity c analyzer on multiple patients.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of tickets determined that there is normal complaint activity for alinity c lactate dehydrogenase (ldh) reagent list number 07p74-30, lot number 83390un20.Trending review determined no trends for falsely elevated results for the product.Return testing was not completed as returns were not available.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.A review of the product labeling concluded that the issue is sufficiently addressed.The search did not show any related potential non-conformances, deviations, or non-conformances.Based on the investigation no product deficiency was identified for the alinity c lactate dehydrogenase (ldh) reagent list number 07p74-30, lot number 83390un20.
|
| |
|