| Model Number |
168-550 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Visual analysis of the returned device identified: flat creases and yellow particles in device outer surface.A leak test was performed which identified no leakage.Fill inspection was performed and identified no blockage in the valve.Based on the device analysis the final assessment is: no issues found related with the manufacturing.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii and exchange from textured to smooth implants.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iii and exchange from textured to smooth implants due to the patient's concern with the product.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|