Product complaint (b)(4).Date sent to the fda: 12/08/2020.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.(b)(4).Sixteen unopened samples of product code pdp259, lot # qhmjrk were received for evaluation.During the visual inspection of thirteen unopened samples, no defects were found on the package.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The device performed without any defect noted and the actual complaint sample was not returned for analysis.Actual: four empty opened foil packets, a paper lid, two empty winding former and a suture piece broken in two sections of product were received for evaluation.During the visual inspection of the sample, the suture pieces were noted damaged and present body fluids due to use.The end of the suture was observed broken due to stress applied that appear to be by use.The product code pdp259 contains an absorbable suture and the time of exposure of suture to the environment could not be determined.Due to condition of the sample received no functional test could be performed.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.Attempts are being made to clarify the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.What is the specific number of devices (total qty involved) suture broken during this procedure? please confirm the specific number of patient events, per suture product and the quantity of sutures involved.What was used to complete the procedure? if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Events were submitted 2210968-2020-09735, 2210968-2020-09737, 2210968-2020-09738, 2210968-2020-09739 and 2210968-2020-09740.
|