• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER UNICEL DXH 900 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER UNICEL DXH 900 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number TN,DXH 900 HEMATOLOGY SYSTEM WITH FLOOR STAND
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The available information reasonably suggests the reported issue was a use error.The fse placed a paper towel underneath the aspiration probe while it was in motion piercing his finger.Bec internal identifier case: (b)(4).
 
Event or Problem Description
A beckman coulter field service engineer (fse) punctured his finger on the piercing probe of a customer's unicel dxh 900 coulter cellular analysis system while servicing it.The fse went to the customer lab site for a leaky dxh900 aspiration probe.While addressing the leak, the fse placed a paper towel underneath the aspiration probe while it was in motion.The fse stated he had done this procedure before and thought he had a few more seconds, the probe extended and pierced his left index finger.The fse went to the emergency room where blood was drawn to screen for (b)(6).Baseline testing was performed; the first blood results came back b)(6).The fse will be scheduling two follow up appointments at 6 weeks and 6 months for his second and third round of (b)(6) shots.The fse reported he is healthy and able to continue with work.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 900 COULTER CELLULAR ANALYSIS SYSTEM
Common Device Name
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd
brea CA 92821
MDR Report Key10973328
Report Number1061932-2020-00176
Device Sequence Number791460
Product Code GKZ
UDI-Device Identifier15099590701574
UDI-Public(01)15099590701574(11)191014
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K140911
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTN,DXH 900 HEMATOLOGY SYSTEM WITH FLOOR STAND
Device Catalogue NumberC11478
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2020
Initial Report FDA Received Date12/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-