BIOMET 3I 3I T3 EX HEX PARALLEL WALLED IMPLANT 4 X 8.5MM; DENTAL IMPLANT Back to Search Results
Model Number BOES485
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Information (3190)
Date of Event 06/09/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).A summary investigation has been completed for lack of primary stability events that do not allege a deficiency with the implant and identified that the reported event is likely due to biological factors which have an adverse effect on implant stability or is related to surgical technique and insertion torque.Due to a wide range of external factors (non-design/ non-manufacturing related), identifying a definitive root cause is generally not possible.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported that the implant at tooth site 29 was stuck in a supra gengival position, there was dense bone and was removed with a implant removal kit instruments, torque 200n.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3 EX HEX PARALLEL WALLED IMPLANT 4 X 8.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key10973841
Report Number0001038806-2020-02026
Device Sequence Number19887265
Product Code DZE
UDI-Device Identifier00844868032502
UDI-Public(01)00844868032502(17)220410(10)2017031648(241)BOES485
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K133049
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/10/2022
Device Model NumberBOES485
Device Catalogue NumberBOES485
Device Lot Number2017031648
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/13/2020
Supplement Date Received by Manufacturer03/26/2021
Initial Report FDA Received Date12/09/2020
Supplement Report FDA Received Date03/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
-
-