| Catalog Number |
2C8931 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event occurred during the unspecified date of (b)(6) 2020 (also reported as "over the course of two days").Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported fifteen (15) clearlink system continu-flo solution sets would not connect.The defect was further described as "the end would not screw together".There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: one device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing, including pressure and clear passage testing was performed with no issues noted.During the luer lock collar verification, it was identified that the collar of the two piece luer lock was not able to turn.The reported condition was verified.The presence of solvent was due to a manufacturing issue.The remainder of the 14 devices were not received for evaluation; therefore, a device analysis could not be completed.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|