| Model Number |
72290105S |
| Medical Device Problem Code |
Electrical Shorting (2926)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.Visual inspection observed no issues.Functional evaluation revealed the unit the does not turn on.The led's on the power supply do illuminate when plugged in. the complaint was verified and is associated with a component failure.Factors, unrelated to the manufacturing and design of the device, which could have contributed to the identified malfunction, include a defective power supply or power entry module or a blown fuse.There are no indications to suggest that the device/product did not meet specifications upon release into distribution.
|
| |
|
Event or Problem Description
|
|
It was reported that the, before a shoulder procedure, the werewolf controller had a smoking smell coming from it.No patient involvement reported.Results of investigation revealed the unit the does not turn on which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|