Model Number 72200873 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference (b)(4).
|
|
Event Description
|
It was reported that, the "dii controller" got burned, the device overheated.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H10: h2, h3, h6.The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of overheating was confirmed.Product failed functional testing with a short circuit error in port a only.Cause of errors is a defective electronic component on the main digital pcb.Product passed functional testing with a known good main pcb installed.The complaint was confirmed and the root cause has been determined to be a defective electronic component.The device has been in service for over 5 years, thus any malfunction presented at this point in time would not be related to the manufacture of the product.A complaint history review concluded this was a repeat issue.
|
|