SMITH & NEPHEW, INC. DII CONTROLLER; ARTHROSCOPE Back to Search Results
Model Number 72200873
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2020
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that, the "dii controller" got burned, the device overheated.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10: h2, h3, h6.The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of overheating was confirmed.Product failed functional testing with a short circuit error in port a only.Cause of errors is a defective electronic component on the main digital pcb.Product passed functional testing with a known good main pcb installed.The complaint was confirmed and the root cause has been determined to be a defective electronic component.The device has been in service for over 5 years, thus any malfunction presented at this point in time would not be related to the manufacture of the product.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DII CONTROLLER
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10979199
MDR Text Key220688539
Report Number1643264-2020-02072
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010607409
UDI-Public03596010607409
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200873
Device Catalogue Number72200873
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2020
Date Manufacturer Received11/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-