• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX; TRACHEOSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX; TRACHEOSTOMY Back to Search Results
Model Number 8.00MM
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/01/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Information received a smiths medical tracheostomy|pvc - portex tubes blue line ultra (blu) would not hold pressure.The customer confirmed in a pre-use check that the product worked properly.He then started to use it with 30 mmhg of air pressure applied to its cuff.However, in 30 minutes or so, the pressure decreased to 7-10 mmhg.As this event was repeatedly observed, he changed the product to another one in one (1) day after starting to use the first one.No patient injury.
 
Additional Manufacturer Narrative
Other, other text: correction: b1 and h1.Device evaluation: one portex tubes blue line ultra (blu) was returned for analysis.The returned sample was visually inspected at 12? to 16? and normal conditions of illumination, no failures were found.The cuff and pilot balloon of the returned sample were inflated using a syringe with air, then the product was immersed in the water in order to detect any leakage.No air bubbles came out of the cuff or from the pilot balloon.The cuff tear occurs during tracheostomy procedure due to contact with sharp edge.He occurrence for this failure condition could be caused by: cuff evacuation unit and fixture for place trachy tubes, fit cuff inflation and deflation test failure by insufficient thf pvc cement, fit cuff inflation and deflation test failure, improper assembly and partial inflation cuff.Detection procedures include 100 percent visual inspection and 100 percent inflation tests after 12 hrs of resting, cuff has inflated and ensures no deflation has occurred.Also by reinflation tests after 6hrs of resting and 100 percent visual inspection as well as pressure testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Common Device Name
TRACHEOSTOMY
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
MDR Report Key10979240
Report Number3012307300-2020-12247
Device Sequence Number4295096
Product Code BTO
UDI-Device Identifier15019315038808
UDI-Public15019315038808
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K030570
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8.00MM
Device Catalogue Number100/870/080
Was Device Available for Evaluation? No
Date Returned to Manufacturer11/20/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/10/2020
Supplement Date Received by Manufacturer01/10/2021
Initial Report FDA Received Date12/09/2020
Supplement Report FDA Received Date02/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-