It was reported that, the renasys touch cannot generate and control negative pressure.Error code 808 f is displayed.Calibration failed twice.No patient harmed, and no injury reported because this was found during service and repair.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
H10.The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.Suction failure resulting in the loss of negative pressure wound therapy will not directly cause or contribute to serious injury or death, even in a clinical setting.Additional conditions must also exist (e.G.Surgical / treatment conditions; patient factors; the unavailability of back-up or alternate therapies for greater than 24 hours) before an injury could possibly occur, and even under those conditions, the risk of patient harm is low.This event is considered not reportable pursuant to 21 cfr part 803.
|