| Model Number |
9438-06 |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that transmitter failed error occurred.Data was evaluated and the allegation was confirmed.The probable cause was determined to be a low transmitter battery.No injury or medical intervention was reported.
|
| |
|
Event or Problem Description
|
|
The product was evaluated.An external visual inspection was performed and passed.Voltage test was performed and failed due to 0v.A review of the share logs was performed and transmitter failed error was found within the investigation window.The allegation was confirmed.The probable cause was determined to be a low transmitter battery.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|