• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PARAGON 28, INC. PHANTOM SMALL BONE INTRAMEDULLARY NAIL SYSTEM; ORTHOPEDIC NAIL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PARAGON 28, INC. PHANTOM SMALL BONE INTRAMEDULLARY NAIL SYSTEM; ORTHOPEDIC NAIL Back to Search Results
Model Number P30-900-100R
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/10/2019
Event Type  malfunction  
Manufacturer Narrative
The device history record was reviewed and met all material specifications with no deviation identified.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.
 
Event Description
It was reported that the patient underwent a surgical procedure that utilized paragon 28 phantom small bone intramedullary nail system on (b)(6) 2019.It was reported that the compression screw was found fully compressed and torque indicating driver tips broke off in the compression slot.The initial reporter mentioned the attempt to reverse the loaded compression however, the compression screw broke off the driver in the process.Another set was used to complete the case.This is report 2 of 2 for the incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PHANTOM SMALL BONE INTRAMEDULLARY NAIL SYSTEM
Type of Device
ORTHOPEDIC NAIL
Manufacturer (Section D)
PARAGON 28, INC.
14445 grasslands dr.
englewood CO 80112
Manufacturer Contact
victoria akinboboye
14445 grasslands dr.
englewood, CO 80112
7206431300
MDR Report Key10981389
MDR Text Key221290054
Report Number3008650117-2020-00241
Device Sequence Number1
Product Code KTW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182307
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberP30-900-100R
Device Catalogue NumberP30-900-100R
Device Lot NumberTN55512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/10/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-