| Model Number |
623-00-40E |
| Medical Device Problem Codes |
Degraded (1153); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Ambulation or Postural Difficulties (2544)
|
| Date of Event |
11/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
As reported: "revision [right] total hip done (b)(6) 2020.Labs done for infection at time of surgery were negative.Implants can not be retrieved due to hospitals policy." revision due to dissociation of the head from the stem.Liner wear noted around it's rim.V40 head, 0° poly liner, and tmzf stem were revised to a 10° 36 mm eccentric liner and competitor femoral components.No further information will be released by the hospital or surgeon.
|
| |
|
Event or Problem Description
|
|
As reported: "revision [right] total hip done (b)(6) 2020.Labs done for infection at time of surgery were negative.Implants can not be retrieved due to hospitals policy." revision due to dissociation of the head from the stem.Liner wear noted around it's rim.V40 head, 0° poly liner, and tmzf stem were revised to a 10° 36 mm eccentric liner and competitor femoral components.No further information will be released by the hospital or surgeon.
|
| |
|
Additional Manufacturer Narrative
|
|
Re[ported event: an event regarding wear involving a trident liner was reported as part of a disassociation event between a lfit v40 cocr head and an accolade stem.The event of disassociation was confirmed through clinician review of the provided medical records.Wear of the liner was not confirmed.Method & results: product evaluation and results- no product was returned for evaluation however, photographs were provided.The photograph show a recently explanted liner.Blood is visible on the liner.Wear is visible on the rim of the liner.Clinician review: a review of the provided medical records by a clinical consultant stated the following comment: confirmation: based on the x-ray and the pi report the event can be confirmed.Root cause: the root cause of the dissociation cannot be stated without additional information.Obtaining the explants may contribute to determining the root cause or evidence/source of material failure.Operative notes or clinical documents particularly intraoperative findings may also contribute.Product history review: not performed as no lot information was provided.Complaint history review: not performed as no lot information was provided.Conclusion: based on the clinician review the event of disassociation was confirmed.Wear was not confirmed.Additional information such as obtaining the explants may contribute to determining the root cause or evidence/source of material failure.Operative notes or clinical documents particularly intraoperative findings may also contribute.No further investigation for this event is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.H3 other text: device not returned.
|
| |
|