Model Number 71441147
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Case reference: (b)(4).
 
Event or Problem Description
It was reported that, a genesis ii mis dcf distal cut blk was starting to rust.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident this device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII MIS DCF DISTAL CUT BLK
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10982783
Report Number1020279-2020-07322
Device Sequence Number4258806
Product Code HRY
UDI-Device Identifier03596010497185
UDI-Public03596010497185
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71441147
Device Catalogue Number71441147
Device Lot Number19HM23965
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/18/2020
Supplement Date Received by Manufacturer05/04/2021
Initial Report FDA Received Date12/10/2020
Supplement Report FDA Received Date05/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-